ISO 10993-4:2017
p
ISO 10993-4:2017
63448

Status : Published (Under review)

This standard was last reviewed and confirmed in 2023. Therefore this version remains current.
en
Format Language
std 1 194 PDF
std 2 194 Paper
  • CHF194
Convert Swiss francs (CHF) to your currency

Abstract

ISO 10993-4:2017 specifies general requirements for evaluating the interactions of medical devices with blood.

It describes

a) a classification of medical devices that are intended for use in contact with blood, based on the intended use and duration of contact as defined in ISO 10993‑1,

b) the fundamental principles governing the evaluation of the interaction of devices with blood,

c) the rationale for structured selection of tests according to specific categories, together with the principles and scientific basis of these tests.

Detailed requirements for testing cannot be specified because of limitations in the knowledge and precision of tests for evaluating interactions of devices with blood. This document describes biological evaluation in general terms and may not necessarily provide sufficient guidance for test methods for a specific device.

The changes in this document do not indicate that testing conducted according to prior versions of this document is invalid. For marketed devices with a history of safe clinical use, additional testing according to this revision is not recommended.

Read sample 

Preview this standard in our Online Browsing Platform (OBP)

General information

  •  : Published
     : 2017-04
    : International Standard confirmed [90.93]
  •  : 3
     : 69
  • ISO/TC 194
    11.100.20 
  • RSS updates

Life cycle

Sustainable Development Goals

This standard contributes to the following Sustainable Development Goal

Got a question?

Check out our FAQs

Customer care
+41 22 749 08 88

Opening hours:
Monday to Friday - 09:00-12:00, 14:00-17:00 (UTC+1)